X-BEAM 300

ON-OFF
Push Switches.
Major
& Minor Power Line Compensation.
Radiographic
KVP Selection from 40 KVP to 100 KVP in steps of 2 KVP.
Technic
Selector – To select 50mA, 100mA, 200mA & 300mA.
Fluoro,
Radiography, Bucky Radiography, Accs selection.
Fluoro
timer reset.
Preparation
& Exposure Push Switches.
LED
indication for Line on, X-Ray on & Tube overload.
Power
Line Compensation Indication.
mA
indication for Radiography & Fluoroscopy.
Solid
State Electronic Timer in 24 steps from shortest exposure time of 0.02 sec
and Maximum 5 sec.
Spot
Film facility at all mA Technics.
The
system will trip off in case of overload in H.T. or L.T. Circuit.
There
will be no exposure if the selection of exposure factor exceeds Tube
Ratings.
X-Ray
initiation by means of Hand Switch or from Control.
Fluoroscopic
Timer.
Technical
Specification :
Full
wave Rectified Two Pulse X-Ray Generator.
Output
– 300mA at 100KVP
Radiographic
Voltage – 40 KVP to 100 KVP in steps.
Fluoroscopic
Voltage – 50 KVP to 100 KVP in steps.
Fluoroscopic
mA – 0 – 3 mA.
X-Ray
Tube – BEL DRA-1/THX-1 Rotating Anode Tube Unit or equivalent.
Overload
– Fully protected.
Single
Tube operation with provision for coupling second tube.
Power Requirement – 400V/50Hz with line resistance .5 ohm to be connected through phase Reducing Transformer.