X-BEAM
60
Key
Features
of
X-BEAM 60
Major
and Minor Line Voltage Compensation.
Exposure
time selection from .02 sec. to 5 sec. in 24 Steps.
KVP
Selection from 45 KVP to 100 KVP in steps of 5 KVP.
Technic
Selector – To select 20mA, 40mA & 60mA.
LED
indication of POWER ON, X-RAY ON, OVERLOAD.
X-Ray
initiation by means of Hand Switch or from Control.
Full
wave Rectified Two Pulse X-Ray Generator.
Tube
overload protector – Incase the selection of KV, mA & Time exceeds the
tube limit, the system will automatically get jammed and there will be no
exposure.
Incase
of overload, entire system will automatically trip off.
High
mobility helps the operators to maneuver the unit easily.
Counter
balance column with dual rope and safety lock as per BIS specification.
Foot
operated breaking system.
Counter
balance vertical movement of Tube Head.
Rotation
of Tube Head around its axis.
Technical
Specification:
Type
of Generator: Two Pulse Full wave Rectified.
X-Ray Tube Unit: Full wave
Rectified with BEL-DSA-3 Stationary Anode Tube Insert. focal spot 2.8mm x
2.8mm.
Collimator: Manual Light
Beam Diaphragm.
Vertical Movement of Tube
Unit – 1192mm.
Horizontal Movement of Tube
Unit – 230mm.
Rotation of Tube Head around
its axis – 270o
Rotation of Tube Head around
the axis of Tube Arm - 3600
Horizontal Swivel of Tube
Head - 1800